Fluke 5020A User Manual
Page 87

Digital Communications Interface
Commands
6
6-17
ALARm:RHUMidity
Sets the humidity rate alarm limit for the specified channel in %RH per hour.
Example: ALAR:RHUM1:RATE:LIM 1
The RHUMidity suffix,
specifies the value for the humidity rate alarm limit. Specifying the MIN, MAX, or DEF
parameter sets the humidity rate alarm limit to the minimum, maximum, or default value
respectively. This command may be password protected (see Password Commands).
ALARm:RHUMidity
Indicates whether a humidity sensor alarm event has occurred for the specified channel.
Example: ALAR:RHUM1:SENS?
Response: 1
The RHUMidity suffix,
humidity sensor alarm event has occurred on the specified channel. A value of 0 is
returned if a humidity sensor alarm event has not occurred on the specified channel.
ALARm:RHUMidity
Returns the state of the humidity sensor alarm enable for the specified channel.
Example: ALAR:RHUM1:SENS:ENAB?
Response: 0
The RHUMidity suffix,
the humidity sensor alarm is enabled. A value of 0 is returned if the humidity sensor
alarm is disabled.
ALARm:RHUMidity
Sets the state of the humidity sensor alarm for the specified channel.
Example: ALAR:RHUM1:SENS:ENAB 1
The RHUMidity suffix,
turns the humidity sensor alarm on (1 or ON) or off (0 or OFF). This command may be
password protected (see Password Commands).
ALARm:RHUMidity
Indicates whether an upper humidity alarm event has occurred for the specified channel.
Example: ALAR:RHUM1:UPP?
Response: 0
The RHUMidity suffix,
an upper humidity alarm event has occurred on the specified channel. A value of 0 is
returned if an upper humidity alarm event has not occurred on the specified channel.
ALARm:RHUMidity
Returns the state of the upper humidity alarm enable for the specified channel.
Example: ALAR:RHUM1:UPP:ENAB?
Response: 0