Measurement commands – Fluke 5020A User Manual
Page 92

5020A
Users Manual
6-22
ALARm:TIME:LAST?
Returns the time of the last alarm event.
Example: ALAR:TIME:LAST?
Response: 15,43,19
The response is returned in the format,
Measurement Commands
The measurement commands are used to read information on the statistical calculations
and for reading the most recent measurement for a specified channel.
CALCulate:AVERage:CLEar
Resets all running statistics for both channels.
Example: CALC1:AVER:CLE
This command may be password protected (see Password Commands).
CALCulate
Returns the dewpoint measurement on the specified channel in degrees C or F.
Example: CALC1:DEWP?
Response: 3.5
The CALCulate suffix,
always in the current temperature units (C or F).
CALCulate
Returns the heat index calculation for the specified channel in degrees C or F.
Example: CALC1:HIND?
Response: 28.4
The CALCulate suffix,
always in the current temperature units (C or F).
CALCulate
Returns the specified statistical calculation for the specified channel.
Example: CALC1:PAR1:AVER4?
Response: 25.186
The CALCulate suffix,
specifies the statistic: average (1), standard deviation (2), minimum (3), maximum (4),
spread (5), maximum rate (7), or alarm count (8). Temperature values are in the current
temperature unit, C of F.
CALCulate[
Resets all of the running statistics for both channels.
This is the same as the CALCulate:AVERage:CLEar command.