beautypg.com

Lightning Audio 7532 User Manual

Page 80

background image

Troubleshooting

Radar Tank Gauge

80

Installation and Operations Manual

A270

custody switch

undef

check position

switch for custody transfer

may be defective

check position of custody

switch

exchange electronics

#

inconsistency between phase

and amplitude evaluation

inconsistent microfactor

inconsistent index mapping

check basic calibration

check mounting calibration

check echo quality

reset history "555"

check stilling pipe

diameter

switch off autocorrection

A272

electronics defect

amplifier

inconsistency in amplification

exchange electronics

W275

electronics defect

factory setting

offset drift of A/D commuter

exchange electronics

W511

no factory

calibration ch1

factory calibration has been

deleted

record new factory

calibration

A512

recording of

mapping please wait

mapping active

wait some seconds until

alarm disappears

W601

linearization ch1

curve not monotone

linearization not

monotonously increasing

correct linearization table

W611

less than 2

linearization points

for channel 1

number of entered

linearization points < 2

correct linearization table

W621

simulation ch. 1 on

simulation mode is active

switch off simulation mode

E641

no usable echo

channel 1

check calibr.

echo lost due to application

conditions of built up on

antenna

check installation;

optimize orientation of

antenna; clean antenna

E651

level in safety

distance - risk of

overspill

level in safety distance

alarm will disappear as

soon as level leaves safety

distance;

A671

linearization ch1 not

complete, not

usable

linearization table is in edit

mode

activate linearization table

W681

current ch1 out of

range

current out of range (3,8 mA

... 21,5 mA)

check calibration and

linearization

Code

Description

Possible cause

Remedy

Table 3: System error messages