Commissioning

Radar Tank Gauge

48

Installation and Operations Manual

To successfully commission a precise measurement to the nearest mm, it is important

you carry out a history reset on first installation after mechanical installation and after

the basic setup of the device (see Page 56). Only after a history reset the mounting cali-
bration

is carried out. Enter the measurement offset as the first point in the dip table for

the mounting calibration. When a value is dipped at a later date, make a second entry

into the dip table, again using the semi-automatic mode. This way, you can easily carry

out a linearization of the measurement.

When configuring the function in "basic setup" (00) please take into account the

following notes:
• Select the functions as described on page 31.
• Some functions can only be used depending on the parameterization of the

instrument. For example, the pipe diameter of a stilling well can only be entered if
"stilling well"

was selected beforehand in the "tank shape" (002) function.

• Certain functions (e.g. starting an interference echo mapping (053)) prompt you to

confirm your data entries. Press

O

or

S

to select "YES" and press

F

to confirm. The

function is now started.

• If you do not press a key during a configurable time period (→ function group

"display (09)"), an automatic return is made to the home position (measured value

display).

Note!
• The instrument continues to measure while data entry is in progress, i.e. the current

measured values are output via the signal outputs in the normal way.

• If the envelope curve mode is active on the display, the measured values are updated

in a slower cycle time. Thus, it is advisable to leave the envelope curve mode after the

measuring point has been optimized.

• If the power supply fails, all preset and parameterized values remain safely stored in

the EEPROM.

Note!

All functions are described in detail, as is the overview of the operating menu itself, in

the 7500 RTG "Service manual and description of instrument functions".