Figure a-2. inx — initiate op code command, Figure a-3. fth — fetch command – KEPCO MAT 1/3 Rack User Manual
Page 47
A-2
MATSVC 112404
FIGURE A-2. INX — Initiate Op Code Command
FIGURE A-3. FTH — Fetch Command
Syntax:
INX VOLT (initiate voltage reading)
INX CURR (initiate current reading)
Function:
Commences a data acquisition process in accordance with the preceding FNC command.
Description:
The response to the INX command is a dynamic time-out value, unless a catastrophic error condition
exists, in which case an error message will be returned. If the time-out value returned is not zero, this
indicates the power supply’s output voltage or current has not yet settled. A time delay should be
observed before proceeding with the FTH command, or the command may be repeated until a zero
value is returned, but the preceding FTH command must also be repeated.
Example:
INX VOLT
Power supply initiates voltage reading)
FTH VOLT
Power supply sends voltage reading to controller)
INX
Syntax:
FTH VOLT (fetch voltage reading)
FTH CURR (fetch current reading)
Function:
Commands the previously designated power supply to return the requested data reading.
Description:
This command must immediately follow an INX command. The value returned is the value of the out-
put voltage or current, whichever was requested, unless a catastrophic error condition exists, in which
case an error message will be returned. The value observed will be in scientific notation.
Example:
INX VOLT
Power supply initiates voltage reading)
FTH VOLT
Power supply sends voltage reading to controller)
FTH