3 terminations, Terminations -2, Internal controls accessible through top cover -2 – KEPCO MAT 1/3 Rack User Manual

Page 19

background image

2-2

MAT 1/3 RACK/ 112404

If a transport cover is not available, the MAT power module may be mounted on a 3/4 inch thick
(or greater) piece of wood by using 8-18 thread-forming screws with a type B style point. A
screw should be used to secure the wood to each of the mounting grommets in the bottom of
the chassis.

2.3

TERMINATIONS

1.

Internal Calibration Controls: Refer to Table 2-2 and Figure 3-2.

2.

Front Panel: Refer to Table 2-3 and Figure 2-1.

3.

Rear Panel: Refer to Table 2-4 and Figure 2-2.

TABLE 2-2. INTERNAL CONTROLS ACCESSIBLE THROUGH TOP COVER

REFERENCE

DESIGNATION

CONTROL

PURPOSE

A1R28

I

O ZERO

(NOTE 1)

Used to adjust output current of power supply to initial value.
CONDITIONS: current mode, shunt at the output.

A1R20

E

O ZERO

(NOTE 1)

Used to adjust output voltage of power supply to zero.
CONDITIONS: voltage mode, no load.

A1R15

I

CURRENT SENSING

ZERO

(NOTE 1)

Used to adjust current monitor amplifier to zero.
CONDITIONS: voltage mode, no load.

A1R16

I

FULL SCALE CURRENT ADJUST

(NOTE 1)

Used to calibrate the full scale value of the output current.
CONDITIONS: current mode, shunt at the output.

A1R11

V

REF

(NOTE 1)

Used to calibrate full scale voltage.
CONDITIONS: voltage mode, no load.

A1R12

I

REF

(NOTE 1)

Used to adjust internal full scale current.
CONDITIONS: current mode, shunt at the output.

A3S1

Node Address

Used to select Node address (node number or channel number) of the
MAT Power Supply from 1 - 31

A1TP1

Test Points

Provides four test points used for testing/calibration
1 -CURRENT REFERENCE
2 -CURRENT MONITOR
3 - VOLTAGE REFERENCE
4 - COMMON

NOTES:

1. These adjustments must be done in accordance with the calibration procedure (see PAR. 5.5).