Change [settings] for gauge operating parameters, Capture [baseline] pressure, Settings – Retrotec DM32 User Manual
Page 32: 1 capture [baseline] pressure

Page 32 of 94
©Retrotec Inc. 2014
4. Change [Settings] for gauge operating parameters
Tapping [Settings] from the Home screen will bring up a menu that lists gauge settings. There are three
“Settings” screens, each displaying a number of gauge operating parameters that can be changed.
Figure 34: [Settings] can be changed on the gauge, and are remembered when the gauge is turned back on
More specifically, the [Settings] allows a user to:
acquire and apply bias pressure correction by tapping
enter the conditioned
(sometimes needed by “Result to be displayed”)
enter the number of seconds for
change the type of, and units for, the “Result to be displayed”
to toggle connection types and edit network credentials
set a [Default @ Pressure] for Result extrapolation when [Set Speed] is used
select the inactivity time (15 min, 30 min, 1 h, 2h, Off), after which the gauge powers itself down,
. The gauge will not power itself down if “Off” is chosen.
enable or disable [Sound] feedback upon key press
set an appropriate “n” value, a slope for result extrapolation, with [Ducts=0.6, House=0.65]
update gauge firmware by tapping
Configurator software – refer to section 6)
Reset touchscreen sensitivity with
For screen illustrations on using a function, click the hyperlinks in the above list.
4.1
Capture [Baseline] Pressure
Measure and remove bias/Baseline pressures due to wind or stack that are present before test pressures
are applied. The captured Baseline is subtracted from the pressure reading on [Channel A].
1. Tap [Capture Baseline] to begin sampling
2. Tap [End Capture] after 30 to 60 seconds. The displayed Baseline pressure is now subtracted from
all future readings on [Channel A]