Output format of the parameter list – HEIDENHAIN ND 221 B v.2 User Manual
Page 36
36
Parameter List for ND 221 B: (factory default setting)
Parameter List
Description
*
Start character (*);
ND-221 B MM
Device; MM or IN;
P01 MM = MM =
0
Unit of measure: MM = 0; INCH = 1;
P11 SCL = SCALING.OFF =
0
SCALING FACTOR OFF = 0; ON = 1;
P12 SCL = 1.000000
SCALING FACTOR = 1.000000; (value input without sign)
P30 DIR = DIRECT. POS =
0
COUNTING DIRECTION POS = 0; NEG = 1;
P31 S. PER. = 10
SIGNAL PERIOD = 10 µm; (value input without sign)
P33 STEP = COUNT 0-5 =
5
COUNTING MODE 0-5 = 5; 0-2 = 2; 0-1 = 1;
P38 DEC = DP POS. 4 =
4
DECIMAL PLACES 4 (range: 1-8)
P40 COMP. = COMP. OFF =
0
COMPENSATION OFF = 0; LIN = 1; MULTI = 2;
P41 L.COMP. = + 0.0
LINEAR COMPENSATION = 0 µm/m (value input)
P42 BKLASH = + 0.0000
BACKLASH compensation = 0.0000 mm (value input)
P43 REF = SINGLE REF. =
0
SINGLE REFERENCE MARK = 0; 500; 1000; 2000; 5000 SP;
P44 REF = REF. ON =
1
REF.ON = 1; REF.OFF = 0;
P45 ALARM = FRQ. + CONT. =
3
OFF= 0; FRQ.= 1; CONT.= 2; FRQ+CONT. = 3;
P50 RS232 = 9600 BAUD = 9600
BAUD RATE = 9600; (110-38400)
P51 RS232 = EMPTY LINE =
1
LINE FEES = 1; (0-99)
P79 PRESET = + 0.0000
DATUM SETTING = 0; (value input)
P80 ENT-CL = CL-ENT OFF =
0
CL-ENT OFF =0; CL-ON = 1; CL...ENT = 2;
P82 DRO.ON = MESSAGE. ON =
1
DISPLAY: MESSAGE.ON = 1; MESSAGE.OFF = 0;
P86 MOD = PRINT OFF =
0
MOD KEY: PRINT OFF = 0; PRINT ON = 1;
P98 LANGUA. = LANGUAGE.DE =
1
NATIONAL LANGUAGE: 0 = EN; 1 = DE; 2 = FR;
*
3 = IT; 4 = NL; 5 = ES;
6 = DA; 7 = SV; 8 = FI;
9 = CS; 10 = PL; 11= HU;
12 = PT;
Final character (*);
Output Format of the Parameter List