List of operating parameters – HEIDENHAIN ND 221 B v.2 User Manual

Page 17

background image

17

Parameter

Settings / Function

P43 REF

Reference marks
One reference mark

SINGLE REF.

Distance-coded with 500 • SP
(SP: signal period)

500 SP

Distance-coded with 1000 • SP
(e.g. for HEIDENHAIN LS ...C)

1000 SP

Distance-coded with 2000 • SP

2000 SP

Distance-coded with 5000 • SP

5000 SP

P44 REF

Reference mark evaluation
Evaluate the reference mark

REF. ON

Do not evaluate the reference
mark

REF. OFF

P45 ALARM

Encoder monitoring
No monitoring

ALARM OFF

Monitor the frequency

FREQUENCY

Monitor contamination

CONTAMINAT.

Contamination + frequency FRQ. + CONT.

P50 RS232

Baud rate
110 / 150 / 300 / 600 / 1200 /

2 400 /

4 800 / 9 600 / 19 200 /

38 400 baud

Parameter

Settings/ Function

P41 L.COMP.

Linear error compensation
– 99 999.9 < P41 < + 99 999.9 [µm/m]
Default setting: 0

Example: Find input value for P41
Displayed measuring length ............. L

a

= 620.000 mm

Actual length (measured, e.g.
with the VM 101 comparator
system from HEIDENHAIN) .............. L

t

= 619.877 mm

Length difference .............................

∆L = L

t

– L

a

= – 123 µm

Compensation factor k (= P41):
k =

∆L / L

a

= – 123 µm / 0.62 m ........ k = – 198.4 [µm/m]

P42 BKLASH

Backlash compensation
Input range (mm):

+9.999 to -9.999

Default setting:

0.000

= no backlash compensation

During a change in direction, the mechanical play between the
encoder and table, the so-called backlash, can cause error.
Positive backlash: The rotary encoder moves before the table,
therefore display value is too high (positive value input).
Negative backlash: The rotary encoder moves after the table,
therefore the display value is too low (negative value input).

List of Operating Parameters