14 Switching the pump on in

subsequent power cycles

Subsequent power-up sequences will jump from the start-up screen to the home

screen.

• The pump runs a power-on test to confirm proper functioning of the memory and

hardware. If a fault is found, an error code is displayed (see 19.2 Error codes on

page 67).

• The pump displays the start-up screen with the Watson-Marlow Pumps logo for

three seconds followed by the home screen.

• Start-up defaults are those in place when the pump was switched off last.

Check that the pump is set to operate as you require it.

The pump is now ready to operate.

All operating parameters may be changed by means of key-presses (see Section 15

Pump operation on page 30).

Power interruption

This pump has an auto restart feature which, when active, will restore

the pump to the operating state it was in when power was lost.

See: 18 Main menu on page 50.

Stop / start power cycles

Do not power up/power down for more than 20 starts per hour, whether manually

or by means of the auto restart facility. We recommend remote control where a high

frequency of power cycles is required.