Watson-Marlow qdos30 Universal and Universal+ Pump User Manual

26

Speed: analogue output (J1) (Universal+ only)

A current analogue signal within the range 4-20mA is

available between the O/P (output) terminal and the GND

terminal. The current is fixed and directly proportional to

the pump speed. 20mA = maximum speed, 4mA = zero

speed.

4-20mA

0/P

GND

1

2

J1

Note: If the mA output is to be used for reading from a multimeter (set on mA), a

250Ω resistor is required in series.

Earth shielding terminals

4.8mm spade terminals are supplied for earth shielding

of cables. The input cable and output cable earths can

be connected to either terminal.

Keep 4-20mA and low voltage signals separate from mains power.

Use separate glanded input cables.