Watson-Marlow qdos30 Universal and Universal+ Pump User Manual

22

12.2 Wiring up the terminal connectors

It is the user’s responsibility to ensure the pump’s safe and reliable operation under

remote and automatic control.

Cable entry to the module is via two watertight cable glands on the module cover.

These may be fitted in place of the sealing plugs which are fitted to the side of the

module cover when the pump is shipped.

The number of glands needed depends on the number of connection cables required
and the convenience of the installer. As standard, two ½” cable glands are provided
with the pump.

Recommended control cable conductors for the terminal blocks: metric = 0.14sq mm

- 2.5sq mm solid and 0.14sq mm - 1.5sq mm stranded. USA = 26AWG - 14AWG solid

and 26AWG - 16AWG stranded. Cable: circular. Max/min outside diameter to ensure

a seal when passed through the standard gland: 9.5mm-12mm. The cable section

must be circular to ensure a seal.

For EMC protection, shielded control cable should be used. The shielding should be

terminated to either of the provided spade connections.

Cables should have a minimum temperature rating of 85C.

Note: Screened, circular control cable up to 12 core must be used. The cable screen

must be earthed at both ends of the cable.

Suitable cables for general-purpose use: 300V with extra premium grade PVC jacket

with good flame and moisture resistance.

Suitable cables for industrial use: 300V extra-rugged polyurethane grade jacket with

resistance to oils, fuels, solvents and water.

For convenience of wiring more than 8 conductors per cable may be awkward to

handle.
• Use a suitable 21mm spanner to unscrew the

sealing plugs.

• Screw in the supplied ½” NPT cable glands

complete with new sealing washers in place

of the plug, ensuring that the retaining nut is

properly seated. Tighten the gland to 2.5Nm

using a suitable 21mm spanner, to ensure a

seal. If a different gland is used, it must be

watertight to IP66.