9 Trouble shooting

9.1 Trouble-shooting

Different kind of problems can cause machine stop, errors etc. Most trouble-shooting will be based on
the information from the front plate indicators and display and the Alarm list information (see 8.1.1)

Examples:

Problem:

Display is flashing AL / 01 when applying power to the module, indication that the
MC100 cannot find any pumps connected.


Solution:

First check if power is applied to the pump(s) and they are ON, then check cabling and
connectors for faults.

To check if problem has been solved, it’s necessary to turn power OFF and ON again


Problem:

Display is flashing AL / 03 when applying power to the module.


Solution:

First check if all the pump(s) has power applied and are indicating power ON.
If this is OK, then the MC100 has stored a different configuration and must be
reconfigured using the factory default configuration.
Make a factory reset to default by following the steps below:

1. Locate the dipswitches on the right side of the cabinet, turn dipswitch 8 ON.
2. Turn power ON to the module, which now starts up with the Alarm, but shortly after

starts the reset sequence and starts up again without the alarm 03.

3. Turn power OFF and turn dipswitch 8 back to OFF.
4. Turn power ON again and confirm that the alarm is still removed.

Alternative:
Use the Node address switch to do a factory reset. see section 5.1

Important:

The cyclic bytes exchanged with fieldbus master/scanner will now reflect
the current configuration. If this configuration is different from setting in
the Master/scanner it will not be possible to connect to the scanner.


Problem:

Display is shoving “S.2.”. As a follow up on the configuration problem above, this will be
the display, when the master/scanner has a different configuration of the number of
cyclic bytes.


Solution:

1. If this is the initial configuration, please use section 6. Configuring the fieldbus

network to the MC100 and the supplied documentation and files on the CD-rom to
setup the network.

2. If this situation occurs after the system has been running normally, but after a

factory reset, focus must be moved to the pumps, where probably one or more
pumps is not communicating on the Flexnet.
Use the description in 6.1 Connecting the MC100 and the pumps for the first
time
, to control the number of pumps that are detected on the Flexnet. When the
problem is solved and the correct pumps corresponding to the correct configuration
is verified, it will be necessary to do a factory reset again.