8.2 Warnings handling

All warnings are cleared automatically, when data within limits are received via FB, or with clear-bit:
MC100 command B1.4

In warning 08-13 by the phrase individual pumps, means “real” connected pumps and pump 0
means the pseudo pump, which holds common parameters in parallel and serial mode

8.2.1 Warnings

Warning no.

Warning description

Action

0

No pending warnings

1

Fill volume out of range

Attempt to set parameter that is
outside limits.
Set new parameter.

2

Tube number is not in table

3

Speed setting is out of range

4

Acceleration setting is out of range

5

Reverse setting is out of range

6

Density setting is out of range

7

Calibration is out of range

8

Common calibration in parallel

Calibration attempt for individual
pump, calibrate pump 0 only.

9

Individual calibration in serial

Calibration attempt for Pump 0,
calibrate pumps individually

10

Attempt to setting common parameters to
individual pump in serial mode

Use pump 0 or clear warning

11

Attempt to calibrate common pump in
Individual mode

Clear by calibrating physical
connected pumps