beautypg.com

Sloan 910 Royal Exposed 900 Hydraulic Series User Manual

Page 7

background image

7

1. PUSH BUTTON LEAKS.

The actuator cartridge has an accumulation of lime or its seals are damaged or worn.
Replace with a new HY-32-A cartridge.

2. THE FLUSHOMETER DOES NOT FLUSH AND A SMALL AMOUNT OF LEAKEAGE IS

VISIBLE BELOW THE VALVE.

A. Foreign material lodged in the cartridge. Remove the cartridge and inspect for foreign

material. Clean under running water.

B. The actuator cartridge has an accumulation of lime or its seals are damaged or worn.

Replace with a new HY-32-A cartridge.

C. Plastic tubing is installed incorrectly. Install plastic tubing correctly (Steps 4 and 6).

ACTUATOR CARTRIDGE REMOVAL

PLASTIC PUSH BUTTON ACTUATOR REMOVAL:

(1) Loosen the setscrew in the button flange and remove the button, flange, and spring from

the actuator body.

(2) Unscrew the cartridge from the actuator body.

NOTE: An automatic check valve in the actuator body allows removal of the

cartridge without turning off the water.

METAL PUSH BUTTON ACTUATOR REMOVAL:

(1) Remove the button or actuator assembly from the wall or fixture.
(2) Disassemble the flange or button assembly from the actuator body.
(3) Unscrew the cartridge from the actuator body.

NOTE: The metal Push Button is designed to be vandal-proof and must be removed

from the wall or fixture for service.

3. THE FLUSHOMETER DOES NOT FLUSH OR FLUSHES ONLY ONCE AND WILL NOT

FLUSH A SECOND TIME WHEN THE BUTTON IS PUSHED.

A. The plunger is lodged in the actuator cartridge or the plunger bypass hole is clogged.

Remove actuator housing and cartridge from the flushometer. Clean under running
water. If cartridge parts are worn, deteriorated or limed up and problem persists after
cleaning, replace with a new HY-83-A cartridge.

B. Plastic tubing is installed incorrectly. Install plastic tubing correctly (Steps 4 and 6).

REMOVAL OF THE ACTUATOR FROM THE FLUSHOMETER:

Turn off water at the control stop. Unscrew the actuator housing coupling nut from the
flushometer. Remove the actuator housing from the flushometer. The tubing connections
can be left intact. Carefully remove the actuator cartridge from the flushometer body to
prevent the actuator from abrupt separation due to expansion of an internal spring. If the
actuator cartridge is lodged in the flushometer body cavity, gently grip the exposed portion of
the cartridge with a channel-lock pliers and rotate back and forth to loosen the o-ring seal.
Carefully separate the actuator housing to reveal the spring and plunger.

4. FLUSHOMETER DOES NOT FUNCTION (NO FLUSH).

A. Control stop or main valve is closed. Open control stop or main valve.
B. Relief valve is worn. Replace performance kit (Royal) or inside parts kit (Regal XL).

5. VOLUME OF WATER IS NOT SUFFICIENT TO SIPHON FIXTURE.

A. Control stop is not open wide enough. Adjust control stop for desired delivery of water

volume.

B. Urinal flushometer parts installed in a closet flushometer. Replace inside urinal

flushometer parts with proper closet flushometer parts.

C. Incorrect dual filtered diaphragm assembly (Royal) or inside parts kit (Regal XL) is

installed in flushometer; for instance, urinal assembly inside a closet flushometer, or low
consumption assembly inside a higher consumption fixture. Determine the flush volume
required by the fixture and replace Royal performance kit or inside parts kit. Use valve
label and markings on fixture for reference.

D. Water supply volume or pressure is inadequate. If no gauges are available to properly

measure supply pressure or volume of water at the flushometer, then remove the relief
valve from the dual filtered fixed bypass diaphragm assembly (Royal) or inside parts kit
(Regal XL), reassemble the flushometer and completely open the control stop. If the
fixture siphons, more water volume is required.

For Royal — Install a higher flushing volume Royal performance kit.

For Regal XL — If a 3.5 gpf inside parts kit is installed in the flushometer, then first

flip the refill head (under the diaphragm) to obtain a 4.5 gpf volume. If this volume is still
inadequate, remove the flow ring from the guide to obtain a 6.5 gpf Kit. If additional flow
is still required, try a low pressure guide kit A-175-A (0301104).

IMPORTANT — LAWS AND REGULATIONS REQUIRING PROHIBIT THE USE OF

HIGHER FLUSHING VOLUMES THAN LISTED ON FIXTURE OR FLUSHOMETER.

If the fixture does not siphon or if a Low Consumption flush is required, steps must
be taken to increase the water supply pressure and/or volume. Contact the fixture
manufacturer for minimum water supply requirements of the fixture.

6. FLUSHOMETER CLOSES OFF IMMEDIATELY.

A. Ruptured or damaged diaphragm. Replace Royal performance kit (Royal) or inside parts

kit (Regal XL).

B.

For Regal XL— An enlarged bypass orifice from corrosion or damage. Replace Inside

parts kit.

7. LENGTH OF FLUSH IS TOO SHORT (SHORT FLUSH).

A.

For Regal XL — The diaphragm Assembly and Guide Assembly are not hand tight.

Screw the two assemblies hand tight.

B.

For Regal XL — An enlarged bypass orifice from corrosion or damage. Replace inside

parts kit.

C. Dual filtered fixed bypass diaphragm assembly (Royal) or inside parts kit (Regal XL) is

damaged. Replace Royal performance kit or inside parts kit.

D. Incorrect dual filtered diaphragm assembly (Royal) or inside parts kit (Regal XL) is

installed in flushometer; for instance, urinal assembly inside a closet flushometer, or low
consumption assembly inside a higher consumption fixture. Determine the flush volume
required by the fixture and replace Royal performance kit or inside parts kit. Use valve
label and markings on fixture for reference.

5. LENGTH OF FLUSH IS TOO LONG (LONG FLUSH) OR CONTINUOUS.

A.

For Royal — Metering bypass hole in diaphragm is clogged. Remove the dual filtered

diaphragm assembly. Remove the primary and secondary filter rings from the diaphragm
and wash under running water. Replace Royal performance kit if cleaning does not
correct the problem.

B.

For Regal XL — Relief Valve (A-19-A) is not seating properly or bypass orifice

is clogged. Disassemble the working parts and wash thoroughly.

NOTE: SIZE OF

THE ORIFICE IN THE BYPASS IS OF UTMOST IMPORTANCE FOR THE PROPER

METERING OF WATER INTO THE UPPER CHAMBER OF THE FLUSHOMETER.

DO NOT ENLARGE OR DAMAGE THIS ORIFICE. REPLACE INSIDE PARTS KIT IF

CLEANING DOES NOT CORRECT PROBLEM.

C. Supply line water pressure has dropped and is not sufficient to close the valve. Close

control stop until pressure is restored.

D. Dual filtered fixed bypass diaphragm assembly (Royal) or inside parts kit (Regal XL) is

damaged. Replace Royal performance kit or inside parts kit.

E. Incorrect dual filtered fixed bypass diaphragm assembly (Royal) or inside parts kit

(Regal XL) is installed in flushometer; for instance, urinal assembly inside a closet
flushometer, or low consumption assembly inside a higher consumption fixture.
Determine the flush volume required by the fixture and replace Royal performance kit or
inside parts kit. Use valve label and markings on fixture for reference.

F. White closet relief valve has been used in a urinal flushometer. Replace closet relief

valve (A-19-AC) with black urinal relief valve (A-19-AU).

G. Inside cover is cracked or damaged. Replace the inside cover (A-71).
H. Conditions in the piping system may contribute to the noise. A degree of high pressure

in the piping may be relieved by adjustments to control stop. Other noises created by
loose pipes, lack of air chambers, inadequate pipe sizes, etc., are problems that must be
discussed with the building engineer.

6. CHATTERING NOISE IS HEARD DURING FLUSH.

A. Inside cover is damaged. Replace inside cover (A-71).
B.

For Regal XL — A-156-A segment diaphragm has been installed upside-down.

Reposition the segment diaphragm properly (see markings on the diaphragm).

When further assistance is required, please contact your

local Sloan Representative or Sloan Technical Support at:

1-888-SLOAN-14 (1-888-756-2614)

Thread the plastic sleeve

onto the stop bonnet until it

is snug. Hand tighten only; do

not use pliers or a wrench.

A

Place the metal control stop cap over the plastic sleeve. Use the palm

of your hand to push or “pop” the cap over the fingers of the Sleeve.

The cap should spin freely on the sleeve.

B

CONTROL

STOP CAP

PLASTIC SLEEVE

BAK-CHEK

®

CONTROL STOP

CONTROL STOP

BONNET

Control Stop Cap Removal (Royal)

Use a large flat screwdriver as a lever to remove the

cap from the control stop. Insert the screwdriver blade

between the bottom edge of the cap and the flat

surface of the control stop body as shown. Push the

screwdriver handle straight back toward the wall to

gently lift the cap. If necessary, work the screwdriver

around the diameter of the cap until you can grasp

the cap and lift it completely off the sleeve. The sleeve should

remain attached to the bonnet of the control stop.

A

Install control stop plug onto the

control stop. Wrench tighten

control stop plug to eliminate

vandalism.

ROYAL FLUSHOMETERS

IMPORTANT: DO NOT INSTALL CAP ONTO SLEEVE UNLESS THE SLEEVE HAS BEEN

THREADED ONTO THE BONNET. IF THE SLEEVE AND CAP ARE ASSEMBLED OFF

THE CONTROL STOP, THE SLEEVE WILL NOT COME APART FROM THE CAP.

CONTROL

STOP PLUG

BAK-CHEK

®

CONTROL STOP

REGAL XL FLUSHOMETERS

8 - INSTALL VANDAL RESISTANT CONTROL STOP CAP

TROUBLESHOOTING