Seaward Sentinel User Manual
Page 5

ELECTRICAL SAFETY TESTER
USER MANUAL
3
10) Flashing high voltage indicator
A flashing red LED indicates dangerous situation during high
voltage output is activated.
11) Data lock function
Front panel could be locked or unlocked by utility setup.
12) PLC remote control
The 9-pin interface provides inputs (START, RESET) and
outputs (TEST, PASS, FAIL).
13) Compact size with multi-capability
The compact size is geared with the Safety Test capabilities
including AC Withstanding test, DC Withstanding test,
Insulation Resistance test and Ground Bond test.
14) R/P Output
The output terminal located on the rear panel.
15) Scanner Interface
16) RS-232 and GPIB Interface
The
RS-232 and GPIB
interface enables remote control operation
and signal processing via a PC.
ELECTRICAL SAFETY TESTER
USER MANUAL
4
2. SPECIFICATION (15℃~35℃ RH≦75%)
1) AC Hi-Pot Specifications
Voltage Range
0.100~5.000kV
Voltage Step
5V/step
Voltage Regulation
(line & load)
1% + 5V
Voltage Accuracy
±1% of reading ±10V (above 500V)
Current Sourcing*
30~40mA(above 500V, maximu test time:
180 sec)
0.10~29.99mA(above 500V, continuous
test)
0.10~10mA(below 500V, continuous test)
Current Limit
0.10~40mA, 0.02mA/step
Current Accuracy
±1% of reading ± 80μA
2) DC Hi-Pot Specifications (for GPT-715A/GPI-735A/745A only)
Voltage Range
0.100~6.000kV
Voltage Step
5V/step
Voltage Regulation
(line & load)
1% + 5V
Voltage Accuracy
±1% of reading ±10V(above 500V)
Current Sourcing*
0.10~10.00mA(above 500V, continuous
test)
0.10~2mA(below 500V, continuous test)
Current Limit
0.10~10mA, 0.01mA/step
Current Accuracy
±1% of reading ± 50μA