beautypg.com

Seaward Sentinel User Manual

Page 12

background image


ELECTRICAL SAFETY TESTER

USER MANUAL



17

Output voltage or current for each step

A C : Output voltage (0.100~ 5.000 kV)

D C : Output voltage (0.100~ 6.000 kV)

I R : Output voltage (50V/100V/500V/1000V)

G B : Output current (3.00~42.00A)

Output
Voltage/Current

C N T : Output current (10mA)

The status of tester includes:
M E N U : Browse and check steps of test.
E D I T : Edit parameters
S A V E : Save parameters
U T I L : Browse and check system utility.
R E A D Y: Ready for test
T E S T : Testing
P A S S : The result of test is pass
F A I L : The result of test is fail

Status

S T O P : Stop the test

ARC

If the ARC function is enable, the sign “*” means that there
is ARC during test.

Lower and upper limit of measurement
I m a x / I m i n : Current measurement limit (ACW & DCW)

Measurement
Limit

R m a x / R m i n : Resistance measurement limit(IR&GB&CNT)
Ramp time and test time
A C : Ramp (000.0~999.9s)

Test (000.5~999.9s)

D C : Ramp (000.0~999.9s)

Test (000.5~999.9s)

I R : Test (001.0~999.9s)
G B : Test (000.5~999.9s)

Ramp/Test
Time

C N T : Test (000.5~999.9s)

time

voltage

ramp

test

time


ELECTRICAL SAFETY TESTER

USER MANUAL



18

CHANNEL
Hi: 00
Lo: 00

Set the Scanner Box to output and return function
through the setting of CH1 and CH2. If the setting is
Hi=01 and Lo=02, that means the high voltage is
output from CH1 and return from CH2. If Hi and Lo
are set to 00, that means no signal is sent to Scanner
Box.

* When the measured load becomes capacitive, the test time must be

extended.

* Ramp Time

Ramp Time setting provides a fast and safe way to detect short circuit

or faulty wiring of your Device Under Test.

GPT-700A monitors the voltage level during the first stage of Ramp

Time to find out short condition. If the voltage still remains less than

80V after 0.1s, GPT-700A cuts off the output and displays the

message “Short”. Therefore the short current will be shut off before it

grows beyond Imax, the current limit setting.

When checking GPT-700A current accuracy, if the DUT has less than

2Mohm input impedance, turn off the Ramp Time setting. GPT-700A

may shut down the voltage taking the low impedance as a sign of

short circuit.