U t i l – Seaward Sentinel User Manual
Page 16
ELECTRICAL SAFETY TESTER
USER MANUAL
25
z To View the System Utility
1. Press the UTILITY key to enter status UTIL.
2. Use the knob or arrow keys to change the active parameter.
G R O U P T E S T
F R O M S T E P
1
UTILITY
U T I L
ELECTRICAL SAFETY TESTER
USER MANUAL
26
3. Table of system utility:
Parameter Option
Description
From STEP 1
The group test procedure always begins
from step 1 to end of group. e.g. 01~01
GROUP TEST
From the
present step
The group test procedure always begins
from the step selected to end of group.
e.g. 01~03
DISABLE
Disable the function arc detection.
ENABLE &
STOP
Enable the arc detection and stop the
test when arc is active.
ARC MODE
ENABLE &
CONTINUE
Enable the arc detection and continue
the test when arc is active.
ARC CURRENT
Set the current level of arc detection.
50 Hz
Set the AC hi-pot output frequency to
50 Hz.
AC
FREQUENCY
60 Hz
Set the AC hi-pot output frequency to
60 Hz.
MODE 1
Control mode of front panel.
Mode 1: Reset first (press reset button
before test)
MODE 2
Mode 2: Press start button directly.
MODE 3
Mode 3: REMOTE I/O enable (the start
button is disable)
TEST
CONTROL
MODE
MODE 4
Mode 4: Reserved.
UNLOCKED
Accept and save all parameters of test
step and utility.
DATA LOCK
LOCKED
Refuse to change any parameters of test
step and utility.
CONTI.
CALIBRATION
Calibrate the short resistance of test
leads for test of Continuity check.