ELECTRICAL SAFETY TESTER

USER MANUAL



35

z Ground Bonding Test (GB- only for GPI-745A)

Press MENU key to enter status MENU then use knob and arrow

keys to select a step.

Press EDIT/SAVE key to enter status EDIT. Now the cursor stays at

the “test mode” field. Use the knob to select mode GB.

Functionality of FIELD key:

Test mode

GB

Output voltage

I=XX.XXA

Measurement upper limit

Rmax= XXX.XmΩ

Measurement lower limit

Rmin=XXX.XmΩ

Testing time

TEST=XXX.Xs

Channel

Hi=00

Lo=00


ELECTRICAL SAFETY TESTER

USER MANUAL



36

Press FIELD key to edit next field “output current”. Use the knob to

adjust the desired output current (3.00~42.00A).

Press FIELD key again to enter next field “measurement upper

limit”. Use the knob to adjust the desired upper limit of resistance

while use the arrow keys to adjust the knob’s resolution
(0~620.0mΩ).

Press FIELD key again to enter next field “measurement lower

limit”. Use the knob to adjust the desired lower limit of resistance

while use the arrow keys to adjust the knob’s resolution
(0~620.0mΩ).

Press FIELD key again to enter next field “testing time”. Use the

knob to adjust the desired testing time while use the arrow keys to

adjust the knob’s resolution (0.5~999.9s).

Press FIELD key again to return the first field “test mode” again.

Press EDIT/SAVE key to save all the parameters.

When the Scanner Box is disconnected, the channel selected menu

would not exist. When set channel to Hi:00 and Lo:10, the test

signal is sent to the instrument only, will not be sent to scanner.