beautypg.com

Seaward Sentinel User Manual

Page 22

background image


ELECTRICAL SAFETY TESTER

USER MANUAL



37

Continuity Check (Cnt)

Press MENU key to enter status MENU then use knob and arrow

keys to select a step.

Press EDIT/SAVE key to enter status EDIT. Now the cursor stays at

the “test mode” field. Use the knob to select mode Cnt. The output

current is fixed to 0.100A.

Functionality of FIELD key:

Press FIELD key to enter next field “measurement upper limit”. Use

the knob to adjust the desired upper limit of resistance while use the
arrow keys to adjust the knob’s resolution (0~1.200Ω).

Press FIELD key again to enter next field “testing time”. Use the

knob to adjust the desired testing time while use the arrow keys to

adjust the knob’s resolution (0.5~999.9s).

Press FIELD key again to return the first field “test mode” again.

Press EDIT/SAVE key to save all the parameters.

Test mode

Cnt


Measurement upper limit

Rmax=XXXXΩ


Testing time

TEST=X.XXXs


ELECTRICAL SAFETY TESTER

USER MANUAL



38

Functionality of remote interface

The remote interface provides two inputs (START and RESET) and three

outputs (TESTING, PASS and FAIL).

RESET:

Connect

“RESET”(pin 1) to “COMMON”(pin 3)

will reset or interrupt this machine (as same as the

functionality of RESET button at front panel).

START:

In status READY, connect “TEST”(pin 2) to

“COMMON”(pin 3) will start the test procedure (as

same as the functionality of START button at front

panel).

TESTING 1,

During test, the “TESTING 1”(pin 4) and

TESTING 2

TESTING 2”(pin 5) will be shorted.

PASS 1, PASS 2

If the result of test is “pass”, the “PASS 1”(pin 6)

and “PASS 2”(pin 7) will be shorted.

FAIL 1, FAIL 2

If the result of test is “fail”, the “FAIL 1”(pin 8) and

“FAIL 2”(pin 9) will be shorted.

*The max. power rating of PASS/FAIL/TESTING outputs:

Apply voltage: +/-300V

Carry current:: +/-100mA

Power dissipation: 400mW

TESTING 2

TESTING 1

FAIL 1

FAIL 2

START
RESET
PASS 1
PASS 2

COMMON