Seaward 250 PrimeTest User Manual
Page 40

PrimeTest 250+
Operating Instructions
38
5 Specifi cation
Earth Continuity
Display Range
0.01ohms to 19.99ohms
Measuring Range 0.05ohms to 19.99ohms
Accuracy
± (5% + 2 digits)
Test current
200mA minimum
Test voltage
6V nominal
Insulation resistance
Display
Range 0.01Mohms to 19.99Mohms
Measuring
Range
0.10Mohms to 19.99Mohms
Accuracy
± (5% + 2 digits)
Test voltage
250V / 500V
Test
current
>1mA
Test current
<2mA into 2k
Substitute Leakage Current
Display Range
0.10mA to 19.99mA
Measuring Range 0.25mA to 19mA
Accuracy
± (5% + 2 digits)
Test voltage
40V rms, 50Hz AC
Test current
<10mA into 2k
Differential Leakage Current
Display Range
0.15mA to 19.99mA
Measuring Range 0.25mA to 19mA
Accuracy
± (5% + 2 digits)
Test voltage
mains voltage, 50Hz AC
Touch Leakage Current
Display Range
0.10mA to 3.5mA
Measuring Range 0.10mA to 1.99mA
Accuracy
± (5% + 2 digits)
Test voltage
mains voltage, 50Hz AC
3-Phase Leakage Current
Display Range
0.10mA to 9.99mA
Measuring Range 0.25mA to 9.99mA
Accuracy
± (5% + 2 digits)
Test voltage
mains voltage, 50Hz AC