PrimeTest 250+

Operating Instructions

33

Mains status

No mains

Earth fault

LN

LE

NE

Flash

Flash

OFF

OFF

Flash

Flash

Whenever reSet is displayed on the LCD press the
reset button on the RCD.

The PrimeTest 250+ will perform a IΔn (30mA) 0°
test, the operating time of the RCD will be shown on
the PrimeTest 250+ LCD. If the test passes then the
PrimeTest 250+ will perform a IΔn (30mA) 180° test. As
long as the tests continue to pass the PrimeTest 250+
will continue to test both 0° and 180° at IΔn (30mA).

3.10

Performing a 3-Phase Sequence

The PrimeTest 250+ can be used to measure R

PE

and earth leakage current on 3 phase equipment
when used with the optional 3 phase adaptor
accessory.

Visually inspect the appliance and mains cord

for signs of damage.

• If the appliance passes a visual inspection

proceed with the electrical tests.

Refer to the documentation supplied with the 3

phase adaptor and connect the adaptor in line

with the 3 phase equipment under test.

Connect the 3 phase adaptor test lead to the

mains input socket 9 (refer to fi gure 2) on the

PrimeTest

250+.

• Connect

the

R

PE

test probe to an exposed

conductive part on the 3 phase equipment.

Press the 3 phase key on the PrimeTest 250+

to being the test.

• If there is a contact voltage present on the

conductive parts of the equipment under test,

the test is terminated and a Volt on test Probe

warning message is shown on the display.