Seaward 250 PrimeTest User Manual
Page 31

PrimeTest 250+
Operating Instructions
29
Note: If the tested cord has a wiring fault, one of
the following enunciators will be illuminated in
place of GOOd
• OPE
n
indicates that either the live or neutral
conductor is broken (open circuit)
• Shor indicates that the live and neutral
conductors are shorted together
• C
r
OS indicates that the live and neutral
conductors are reversed
3.5 Testing an extension lead
• Visually inspect the mains cord and plug for
signs of damage.
•
Check that the correct fuse is fi tted.
•
If the cord passes a visual inspection proceed
with the electrical tests.
•
Plug the supplied 0.5m IEC lead into the IEC
socket and into a mains outlet on the extension
lead. Plug the mains plug of the extension lead
into the front panel mains socket on the
PrimeTest
250+.
•
The extension lead can now be tested in the
same manner as an IEC as described above.
- AGL-5 (2 pages)
- Apollo 500 (21 pages)
- B433R (16 pages)
- Checkbox 17 Plus (4 pages)
- Apollo 600 (32 pages)
- B255 (40 pages)
- A203ES (10 pages)
- A252 (35 pages)
- DO4000 (22 pages)
- DO5000 Interface (41 pages)
- DO5000 Series (36 pages)
- DO7e (24 pages)
- DP6 (22 pages)
- DO7010 (44 pages)
- ARE 1557 (31 pages)
- Europa Pac Plus (53 pages)
- Supernova Elite (118 pages)
- HAL Scan (164 pages)
- G2000 (44 pages)
- IRT 1557 (21 pages)
- KD1E (22 pages)
- LLT (2 pages)
- PAC3760 (2 pages)
- PAC 1500Xi (12 pages)
- PAC 1000 (10 pages)
- PAT2000i (29 pages)
- PATGuard Elite Quick start (2 pages)
- PATGuard Elite SQL Quick start (2 pages)
- PATGuard Pro 2 (77 pages)
- PATguard Pro 2 Quick start (2 pages)
- PATGuard Time Manager (12 pages)
- PATGuard Work About Quick start (4 pages)
- PH3 LLT (6 pages)
- PATGuard Work About (53 pages)
- Solar Power Clamp (31 pages)
- PowerGuard Pro (29 pages)
- PowerPlus 1557 (62 pages)
- PowerPlus PC (44 pages)
- 100 PrimeTest (24 pages)
- PATGuard 3 (153 pages)
- 200 PrimeTest (34 pages)
- 220 PrimeTest (34 pages)
- 50 PrimeTest (20 pages)
- RH9A Series (16 pages)