Seaward 250 PrimeTest User Manual
Page 29

•
If a touch leakage test is performed then the
PrimeTest
250+ will prompt the user to start
the Leakage test by fl ashing the Ileak icon.
If no key is pressed the PrimeTest 250+ will
timeout and return to the start screen.
Note: If the LKGE key is pressed momentarily
the touch leakage test duration is 5s.
If the LKGE key is pressed and held for
2s the touch leakage test duration is
2
minutes.
When the duration is set to 2 minutes,
the test can be terminated at any time
by pressing the LKGE key.
• When the LKGE key is pressed to start the
touch leakage test the PrimeTest 250+ will
check the incoming mains supply to ensure
that the voltages are correct. The enunciators
are displayed if there is an error with the
connected mains supply:
•
If the mains voltage is correct, the PrimeTest
250+ will proceed with the touch leakage test.
•
If the touch leakage is less than the factory set
limit a tick is placed next to the Ileak enunciator.
• If all of the tests within the sequence have
passed then the PASS enunciator is illuminated.
PrimeTest 250+
Operating Instructions
27
Mains status
No mains
Earth fault or Live
and Neutral reverse
LN
LE
NE
Flash
Flash
OFF
OFF
Flash
Flash