If a touch leakage test is performed then the

PrimeTest

250+ will prompt the user to start

the Leakage test by fl ashing the Ileak icon.

If no key is pressed the PrimeTest 250+ will

timeout and return to the start screen.

Note: If the LKGE key is pressed momentarily

the touch leakage test duration is 5s.

If the LKGE key is pressed and held for

2s the touch leakage test duration is

2

minutes.

When the duration is set to 2 minutes,

the test can be terminated at any time

by pressing the LKGE key.

• When the LKGE key is pressed to start the

touch leakage test the PrimeTest 250+ will

check the incoming mains supply to ensure

that the voltages are correct. The enunciators

are displayed if there is an error with the

connected mains supply:

If the mains voltage is correct, the PrimeTest

250+ will proceed with the touch leakage test.

If the touch leakage is less than the factory set

limit a tick is placed next to the Ileak enunciator.

• If all of the tests within the sequence have

passed then the PASS enunciator is illuminated.

PrimeTest 250+

Operating Instructions

27

Mains status

No mains

Earth fault or Live
and Neutral reverse

LN

LE

NE

Flash

Flash

OFF

OFF

Flash

Flash