Substitute Leakage Test
43
Substitute Leakage Test
Test Voltage * ................................................................... 40V a.c. o/c
Display Range .......................................................... 0.2 - 20.0 mA a.c.
Resolution. ................................................................................0.01mA
Accuracy...................... +/- 10% of reading, +/- 2 digits (1.00 - 20mA)
Pass Levels. ....... 0.25mA, 0.5mA, 3.5mA, 7mA, 15mA, User defined
Differential Leakage
Range...........................................................................0.0mA - 10.0mA
Resolution.. ...............................................................................0.01mA
Accuracy.......................... +/- 5% of reading, +/- 4 digits (0.2 – 20mA)
Pass Levels ........... 0.25mA, 0.25mA, 0.75mA, 3.5mA, User defined
Touch Leakage
Range.......................................................................... ..0.0mA – 3.5mA
Resolution. ................................................................................0.01mA
Accuracy......................... +/- 5% of reading, +/- 4 digits (0.1 – 3.5mA)
Pass Levels. .......................... 0.25mA, 0.75mA, 3.5mA, User defined
Load Test
Socket .............................................................................230V BS 1363
Measured Voltage.................................... 195 - 300V (Indication only)
Measured Current................................... 0.50 – 16A (Indication only)
Measured Load .............................................0 - 4kVA (Indication only)
External Leakage
Range....................................................................... .0.0mA – 19.99mA
Resolution.. ...............................................................................0.01mA