External Leakage Test
31
Test Duration
The test is continuous and is stopped by either pressing the
START/STOP button or by switching to an un-powered test. In both
cases the test will be stopped and the power to the appliance
removed. Switching to another powered test will stop the previous
test and start the new test without interrupting power to the appliance.
User Selectable Settings
There are no user selectable settings.
External Leakage Test
Switch Position
------------
External Leakage (Clamp)
Test Icon Displayed ---
Warning
Mains voltage applied to appliance
Note
This test requires the use of a current clamp, which is an optional
accessory (see Chapter 5 Accessories).
All Appliances
Plug the current clamp into the appropriate socket. Open the clamp
jaws and clamp around the cable you wish to measure current flow
through. Power is applied to the EUT at the mains socket.
Test Description
The tester will display the current detected by the external clamp. If
this is around a protective conductor, earth leakage will be displayed.
If the clamp is around the phase and neutral conductors then