Seaward Europa Pac Plus User Manual
Page 42

Powered Tests
36
Err 03 with flashing cross
Internal relay fault
Return to your Service Agent if problem persists.
Err 04 with flashing cross
Internal relay fault
Return to your Service Agent if problem persists.
Err 05 with flashing cross
Internal relay fault
Return to your Service Agent if problem persists.
Err 06 with flashing cross
Internal relay fault
Return to your Service Agent if problem persists.
Err 07 with flashing cross
Memory corruption.
Return to your Service Agent.
Powered Tests
The Tester performs a number of checks prior to and during powered
tests. The following primary display messages can occur.
Lo Po
Low power caused by load less than 60w. The user can override this
message if low load is expected by pressing the START/STOP
button.
Hi Cu
High current is expected to be drawn by the EUT. This message
appears if the current is greater than 16A. The user can override this
message, provided the current is not expected to exceed 50A, if it is
caused by a high start up current by pressing the START/STOP
button.