Sensor Alignment -

by Alignment Tool

OPS

Series Encoders, Side and Top Mount Configurations

2.

Alignment

Connect the USB cable to the Alignment Tool with power from either the Universal Power supply or computer

USB port.

The Alignment Tool Signal Level and Sensor head LEDs are now used for alignment. Proper alignment is

indicated by green sensor head and Alignment Tool Signal Level LEDs.

3.

Position the sensor over a section of the scale. Adjust the sensor's Y or

q

z directions until the Signal Level and

sensor head LEDs are green.

4.

Move the sensor across the entire length of travel. The LED should be green over the entire length of travel. If

the LED is Yellow or Red, adjust Y and

q

z and refer to the interface drawing to ensure proper mechanical

design. (Passing over an Index will cause the Alignment Tool Index/Limit LED to flash green.)

Green = Optimal

Performance

Alignment Tool

To align the

sensor move it

in the Y or qz

directions

Z

Y

X

q

z

Sensor

Indicator LED

OPS Side Mount

OPS Top Mount

Yellow = Marginal

Performance

Red = Improper

Performance

NOTE:

If you are having any difficulties getting a green light over

the entire travel length, refer to the OPS interface drawing

to check mechanical design. Also, check to make sure the

scale is properly installed, the sensor is properly oriented

with respect to the scale, and the blue protective film has

been removed from the tape scales.