System throughput, Settling time – Measurement Computing PCI-DAS6023 User Manual
Page 32

PCI-DAS6023 and PCI-DAS6025 User's Guide
Specifications
32
Table 3. Absolute accuracy components - all values are (±)
Range
% of
Reading
Offset
(mV)
Averaged Noise +
Quantization (mV)
1
Temp Drift
(%/ C)
Absolute Accuracy
at FS (mV)
±10 V
0.0558
2.930
1.262
0.0010
9.77
±5 V
0.0563
1.460
0.605
0.0010
4.88
±500 mV
0.1046
0.150
0.057
0.0010
0.73
±50 mV
0.1340
0.024
0.007
0.0010
0.098
1
Averaged measurements assume dithering and averaging of 100 single-channel readings.
Table 4. Differential non-linearity specifications
All ranges
±0.5 LSB typ
±1.0 LSB max
System throughput
Table 5. System throughput specifications
Condition
Calibration Coefficients
ADC Rate (max)
1.
1. Single channel, single input range.
Per specified range
200 kS/s
2.
2. Multiple channel, single input range Per specified range
200 kS/s
3.
3. Single channel, multiple input
ranges.
Default to value for cbAInScan() range
200 kS/s
Note 1:
For conditions 1-2 above, specified accuracy is maintained at rated throughput. Condition 3
applies a calibration coefficient which corresponds to the range value selected in
cbAInScan()
.
This coefficient remains unchanged throughout the scan. Increased settling times may occur
during gain-switching operations.
Settling time
Settling time is defined here as the time required for a channel to settle to within a specified accuracy in
response to a full-scale (FS) step. Two channels are scanned at a specified rate. A –FS DC signal is presented to
channel 1; a +FS DC signal is presented to channel 0.
Table 6. Settling time specifications
Condition
Range
Accuracy
±0.012%
(±0.5 LSB)
±0.024%
(±1.0 LSB)
±0.098%
(±4.0 LSB)
Same range to same range
±10 V
5 µS typ
5 µS max
5 µS max
±5 V
5 µS typ
5 µS max
5 µS max
±500 mV
5 µS typ
5 µS max
5 µS max
±50 mV
5 µS typ
5 µS max
5 µS max
Any range to any range
-
25 µS typ
20 µS typ
15 µS typ