Onsite Qualification Protocols and Records: Operational Qualification
8
© Copyright K-Patents 2011. All rights reserved.
47
8.5 System qualification
Check that the system is the same as defined in the IQ, with no changes.
Definition of requirements
All system equipments remain the same as for the IQ and ancillary equipment IQ is valid.
Date
Signature
Pass
Conditional Pass
Fail
Conditional pass:
8.6 Setting up the system components and devices
Description of requirements
The Pharma Refractometer system comprised of Refractometer sensor, Laboratory Test
Cuvette LTC and Indicating transmitter, is assembled and mounted correctly as described
in the
Section 3 (this document). Initial startup cheks for the Refractometer have been
made according to PROCESS REFRACTOMETER PR-23 INSTRUCTION MANUAL,
Section 5. Also the ancillary thermostatic bath and the sample delivery system (if required)
are connected in accordance with the
Section 3 (this document). The ancillary equipment
is switched on and functional cheks are made in accordance with the corresponding
operating manuals.
Requirement acceptance values
The system and devices are complete and have been set up in compliance with
the instructions.
Failure to meet any of the acceptance values
Terminate the OQ.
Date
Signature
Pass
Conditional Pass
Fail
Conditional pass: