Onsite Qualification Protocols and Records: Installation Qualification

7

© Copyright K-Patents 2011. All rights reserved.

39

7.5 Operating environment

Description of requirements
Ensuring that the environmental requirements for the system regarding temperature
are within specification. Ensuring that the appropriate power supply and power switch
are available.

Requirement acceptance values
Ambient temperature is between 20 and 30 °C (68 - 86°F). An electrical power supply with a
voltage and frequency of 100-230 VAC/50-60 Hz (Optional 24 VDC). A computer (PC, Mac,
PDA or mainframe). An available water supply. Optional: Fume hood if required.

Failure to meet any off the acceptance values
A new environment must be established and the qualification performed again from

Section

7.3.1 (of this document) onwards.

Date

Signature

Pass

*

Conditional Pass

Fail

*Conditional pass:

7.6 Installation

Requirement description
The authorized operator who in accordance with

Section 7.2 (of this document), must read

the start-up and operation instructions in

Section 3 (of this document).

Requirement acceptance values
The relevant sections have been read.

Date

Signature

Pass

*

Conditional Pass

Fail

*Conditional pass: