beautypg.com

K-Patents PR-23 for K-Patents User Manual

Page 27

background image

© Copyright K-Patents 2011. All rights reserved.

27

5

Use of Laboratory Test Cuvette LTC:

Continuous sampling

The Laboratory Test Cuvette LTC for continuous sampling can be used for further testing
of samples in pilot scale-up trials for instance in drug formulation or process scale-up. The
typical size of such laboratory batch is 3-10 liters (100 - 300 fl oz).of a liquid.

With continuous sampling the liquid sample is circulated by an external sample pump so the
LTC mixer is not required. Liquid line hoses are used to connect the external sample pump
to the continuous sampling chamber inlet and outlet.

5.1 Assembly of continuous sampling Laboratory Test

Cuvette

The Laboratory Test Cuvette LTC for continuous sampling is assembled using the same
components as the LTC for static sampling, except for the replacement of the static sampling
chamber with the continuous sampling chamber (Figure 5.1). The continuous sampling
chamber has two pipes at the top for inlet and outlet hose connections to an external
sample pump.

1

2

Continuous sampling

chamber

Compression

spring

Temperature control

fluid - inlet

Temperature control

fluid - outlet

O-ring

21x2 EPDM

O-ring

20.2x3 EPDM

Outer casing

Grove for

O-ring

Figure 5.1 K-Patents Laboratory Cuvette LTC for continuous sampling: Components and assembly