Use of PR-23 Pharma Refractometer with Laboratory Test Cuvette LTC: Static Sampling

3

© Copyright K-Patents 2011. All rights reserved.

21

3.4 Mounting Laboratory Test Cuvette to Sensor

The test cuvette is mounted on the refractometer sensor using a 2.5” EPDM sanitary gasket
and clamp. Position the gasket in its seating groove with the lip downward. Now position
the assembled test cuvette with the gasket groove correctly seated on to the gasket (Figure
3.2).
Attach clamp and lock. Check that the clamp is tight, and if not, adjust the tension using
the clamp adjustment screw.

3

4

Seat seal in

flange groves

Cuvette

EPDM sanitary

gasket

Refractometer

prism

Clamp

adjustment

screw

Figure 3.2 Attaching Laboratory Test Cuvette LTC to Refractometer.