beautypg.com

5 connecting hoses from thermostatic bath – K-Patents PR-23 for K-Patents User Manual

Page 22

background image

Document/Revision No. IM-GB-PR23-ACP: Rev. 1.0 Effective: December 15, 2011

22

Appendix to the Instructions Manual PR-23

3.5 Connecting hoses from thermostatic bath

The temperature control Water Jacket is connected to the Thermostatic bath using the
8 mm (0.3 inch) hoses. These hoses should each have a ball valve inserted between the
thermostatic bath and the water jacket (see Figure 3.3). These valves are to enable the
shutting off of the water flow to the water jacket to allow disassembly for cleaning. The water
jacket inlet pipe is the one positioned nearest the base and the outlet is positioned at the
top. Thus the water circulates through the water jacket from bottom to top, therefore avoiding
air pocket formation within the water jacket.

5

Attach 8mm ID hoses from

Termostatic Bath to Water

Jacket with a Cut-off Valve

set into both the feed and

return hoses

Cuvette outlet

Cut-off valves

Cuvette inlet

Thermostatic water

feed

Thermostatic water

return

Figure 3.3 Attaching hoses to connect LaboratoryTest Cuvette LTC to Thermostatic bath.

Operational procedure checking the operation of Laboratory Cuvette sample mixer: Check
that the magnetically driven mixer paddle starts to rotate when the mixer driver’s power is
switched on (Figure 3.4).

Mixer driver

Magnetically driven

mixer paddle

Figure 3.4 Mixer components.