Preparing the Straumann

®

Anatomic IPS e.max

®

Abutment

The Straumann Anatomic IPS e.max Abutment is prepared according to the step-by-step instructions.

For the selection of the grinding instruments, please refer to the Ivoclar Vivadent Flow Chart

“Recommended grinding tools for IPS e.max zirconium oxide”.

Conduct a Regeneration firing after preparing the abutment.

Individualized Straumann

®

Anatomic IPS e.max

®

Abutment on the model

Please observe the following procedure for conducting the Regeneration firing:

– Clean and dry the individualized abutment with the steam jet.

– Position the abutment on a metal pin on a honey-combed tray.

– Conduct the Regeneration firing in a ceramic furnace (e.g. Programat P700) using the respective parameters.

– Do not adjust the abutment by grinding after the Regeneration firing.

– Do not blast the framework with Al

2

O

3

or glass polishing beads before veneering, since this may cause lasting damage

to the surface.

– Before veneering, clean the abutment with running water or the steam jet.

Firing parameters for the Regeneration firing

Furnace

P300
P500
P700

EP 3000
EP 5000

P80

P100
P200

EP 600

PX1

B

°C/°F

403
757

403
757

403
757

S

min

0:18

0:18

0:30

t

1

°C/°F/min

65

117

65

117

65

117

T

1

°C/°F

1050
1922

1050
1922

1050
1922

H

1

min

15:00

15:00

15:00

L

°C/°F

750

1382

750

1382

750

1382

12 min

t

°C/°F/min

25
45

––

––