10 appendix c - digitiser specifications – Guralp Systems CMG-3ESPCD User Manual
Page 96
CMG-3ESPCD Digital Broadband Seismometer
Appendix C - Digitiser specifications
10 Appendix C - Digitiser specifications
Seismic channels
Number
4
Format
32 bit
Inputs
Differential with transient
protection; ±10 V input
range
Input impedance
1 MΩ, 10 nF
Common mode rejection 110 dB at 10 Hz
Mux channels
Number
8
Format
24 bit
Inputs
Single ended; ±10 V input
range
GPS
Unit
External GPS (CMG-GPS2)
Power
Via GPS connector
Time format
NMEA
Max. cable length
50 m
Sensor control
Functions available
Lock, unlock, and centre
sensor; enable calibration
Sensor calibration
Types
Sine and square wave,
adjustable amplitude and
frequency; broadband noise,
adjustable amplitude
Digital signal
processing
Type
Texas TMS320CV33
Hardware sampling rate 512 KHz
Sample output rates
6 cascaded rates, each 2, 4, or
5 times lower than the
previous rate
Anti-aliasing filters
2 pole
Low-pass filters
FIR
Out-of-band rejection
140 dB
In-band ripple
–140 dB
Trigger modes
STA/LTA, level, external
Digitiser
performance
Output format
32 bit
Noise-free resolution
(NPR)
22.5 bits at 20 samples/s
21.5 bits at 100 samples/s
96
Issue B - November 2013