Guralp Systems CMG-3ESPCD User Manual
Page 88

CMG-3ESPCD Digital Broadband Seismometer
Inside the 3ESPCD
Check that there is two-way communication with the digitiser by pressing
Enter. The digitiser should reply with ok on a new line. Type reboot to
reinitialize the digitiser, and confirm with y. As it is restarting, the digitiser
will report its status over the terminal connection, followed by a maintenance
menu:
I/OPort =$42 : RTModule no I2C ACK @ $78 MPE ARM ANS ROM
PowerForth v6.30
ARM Serial BootStrap v1.100, 11 August 2003
Copyright (c) 2002-3 GSL, EDSL & MicroProcessor Engineering Ltd.
Port 0 38400 baud Port 1 38400 baud Port 2 38400 baud
Guralp Systems Ltd - ARM-BOOT v2.0 mgs 27/04/11 (Build 13)
System Code versions loaded :-
Current 0108:0000 Guralp Systems Ltd - DM+FW v.106 mgs
05/03/12 (Build 57d)
Previous 0110:0000 Guralp Systems Ltd - DM+FW v.106 mgs
05/03/12 (Build 57d)
Backup 0118:0000 Guralp Systems Ltd - DM+FW v.106 mgs
05/03/12 (Build 57d)
DSP Code :
0106:0000 dsp1090.bin Default
0107:0000 dsp1090.bin
Command keys:
C - set real time Clock (2013 11 13 17:17:55 )
I - view/upload InfoBlock
F - run the Forth monitor
S - update System program
O - select Other system program
B - update Boot program
D - update DSP code
T - Toggle default DSP code
Q - Quit maintenance system
5 seconds to auto-start Enter command:
If you do not press a key in the next five seconds, the digitiser will start up
normally. The digitiser has three firmware components, which can be
updated separately: the system program, the DSP code and the boot loader.
7.2.3.1 Loading a new system program
To update the system firmware, press S at the maintenance menu. The
digitiser will then request a transfer using the Xmodem protocol. If you are
using Scream!, a file browser window will appear automatically.
88