CMG-3ESPCD Digital Broadband Seismometer
Calibrating the 3ESPCD
3. Repeat this measurement for several tilt angles.
4. For the vertical sensor, the input acceleration is given by
a = g sin φ
whilst for the horizontal sensor, it is
a = g ( 1 – cos )
φ
Calculate the input acceleration for each of the tilt angles used, and
plot a graph of mass position output against input acceleration.
5. The gradient of the line obtained gives the sensitivity of the coil (in
V/ms
-2
, if g was measured in ms
-2
and the mass position in V.)
6. The coil constant K is equal to this sensitivity divided by the value of
the displacement feedback resistor, given on the calibration sheet.