Fluke Biomedical 956A-201-M2 User Manual
Page 91

A-3
Table A-7. Connector P6 – Auxiliary I/O, 956A-201-M2
Pin Signal
1
4-20 mA Output #1 (+)
2
4-20 mA Ground (-)
3
4-20 mA Output #2 (+)
4
4-20 mA Ground #2 (-)
5
0 – 10 V * (+)
6 Ground
7
Analog Output Option (+), 0-10 mV
8 Ground
(-)
9 through 24
Not Used
* 0 – 10 Vdc selected for Remote Meter use on P2 (10, 11) and is not available on P6 (5, 6).
Table A-8
RS232C Option
Pin Signal
1 Protective
Ground
2 Transmit
Data
3 Receive
Data
4
Request to Send (RTS)
5
Clear to Send (CTS)
6
Data Set Ready (DSR)
7 Signal
Ground
8
Receive Line Signal Detect (DCD)
20
Data Terminal Ready
Table A-9
Model 956A-201 Option Jumpers
Jumper Function
Position
Operation
JP1 Microprocessor
Reset Out
IN
Normal Operation (factory) Momentary connect to
reset
Not Applicable
JP2 PROM
Type
1-2
2-3
PROM 27128
PROM 27256 (factory)
JP3-1/JP3-2
Accuracy (it should be noted
that this accuracy refers only to
the statistical interpretation of
detector counts, and not
detector accuracy)
Out/Out
Out/In
In/Out
In/In
2% Accuracy – 9604 Target Counts, Not implemented
5% Accuracy – 1537 Target Counts, Not implemented
10% Accuracy – 384 Target Counts, Not implemented
Fixed one second display update of 60 one second
values (factory)
JP3-3 Alarm
Acknowledge
In
Out
Manual Acknowledge (factory)
Automatic
JP3-4
Fail Alarm
In
Out
Enable No Counts Fail Alarm (factory)
Disable No Counts Fail Alarm
JP3-5
Check Source Alarm
In
Out
Alarm Inhibited (factory)
Alarm Enabled
JP4 Input
Pulse
Selection 1-2
2-3
Negative Input Pulse
Positive Input Pulse (factory)
JP5 Shield
Polarity
Selection
1-2
2-3
Shield for Negative Pulse
Shield for Positive Pulse (factory)
JP-6 Anti-Jam
Fuse
Selection 1-2
2-3
Enable for Normal Operation (factory)
Disabled for testing only
JP7
Detector Type for Anti-Jam
Circuit Timing
1-2
2-3
Out
Scintillation
G-M Tube (factory)
Anti-jam circuit disabled (test purposes only)