4 specification, detector preamplifier (intergral) – Fluke Biomedical 956A-201-M2 User Manual
Page 8

1-4
1.3 SPECIFICATIONS, G-M DETECTOR CHECK SOURCE,
MODEL 897A SERIES
Radionuclide
36
CI
Source Activity:
8 microcuries
Scale Reading (Approximate)
897A-210,
897A-211
897A-220,
897A-221
897A-230,
897A-231
10-150 mR/h, Typical
10-50 mR/h, Typical
10-50 mR/h, Typical
Voltage
12 – 15 Vdc
Encapsulation
Brass, high temperature soldered
1.4 SPECIFICATION, DETECTOR PREAMPLIFIER
(INTERGRAL)
Input Impedance
>100 k-ohms
Output Impedance
50 ohms
Output Pulse Polarity
Positive
Output Signal (50 ohm terminated)
+ 5 Vdc square-wave
Low Voltage*
+ 15 Vdc (Optional + 10 Vdc is Jumper Selectable)
Power Requirements
+ 15 Vdc @ 120 mA
Maximum Cable Length
2000 feet
Electronic Exposure Life
Approximately 10
5
rads
Discriminator Level
Adjustable from 0 to 1.5 Vdc (nominal value =1.00
Vdc)
Anti-Jam Level
Adjustable from 0 to + 3 Vdc
Anti-Jam Oscillator Frequency
Approximately 50 kHz
Pre-divide* Jumper
Selectable
Analog & Digital Monitor Configuration
Divide by 2, Square wave output
Optional Monitor Configuration
Raw pulse output
* Preamplifier low voltage and pre-divide are selected with jumpers. Refer to the preamplifier circuit
description for more information. Standard Configuration is +15 Vdc Divide by 2.