10 auxiliary equipment – Fluke Biomedical 956A-201-M2 User Manual
Page 14

1-10
Detector Input
Digital Pulse, Up to 2000 feet from UDR, 50 ohm input
impedance
UDR Electronic Accuracy
± 1 digit (± 1% of the displayed value, exclusive of the
detector energy response
Dimensions (H x W x D)
3.5 in. x 5.6 in. x 13.5 in. (8.9 cm x 14.2 cm x 34.3 cm)
Weight
Approximately 3.7 lb. (1.67 kg)
Power
120 Vac
± 10%, 50/60 Hz, 28 watts (240 Vac optional)
Operating Temperature
32°F to 122°F (0°C to + 50°C)
Relative Humidity
0 to 99% non-condensing
Compatible Detector
897A Series, G-M (Geiger Mueller tube)
Heat Loading
Approximately 96 BTU/hr
1.10 AUXILIARY EQUIPMENT
Auxiliary Equipment
Model Description
Model 848-8
Field Calibrator (100 mCi)
Model 848-8A
Field Calibrator (10 mCi)
Model 848-8B
Field Calibrator (20 mCi)
Model 848-8-105
897A Adapter (all types)
Model 848-8-400
897A Stainless Steel Detector Adapter, 897A-
2x1
Model 942TS
UDR Test Set
Model 958A-40
Local Alarm, without meter
Model 958A-10, -20, -30
Local Alarm, with meter
Model 948-1
Rack Chassis, 3 Channel
Model 948A-2
Blank Panel
Model 948-3
Table Top Enclosure
Model 948-10
Panel Adapter
Model 50-100
Interconnecting Cable
Model S157033A2
Line Filter/Surge Suppressor
Model 948B-1A-5
Dual Channel Rack Chassis