Fluke Biomedical 956A-201-M2 User Manual
Page 25

2-7
Table 2-12. Model 956A-201-M2 Option Jumpers
Jumper Function
Position Operation
JP1 Microprocessor
Reset
Out
In
Normal Operation (factory) Momentary
connect to reset microprocessor without
cycling AC power
Not Applicable
JP2 PROM
Type
1-2
2-3
PROM 27128
PROM 27256 (factory)
JP3-1/JP3-
2
Statistical Accuracy (Note: This
accuracy refers only to the
statistical interpretation of
detector counts, and not
detector accuracy)
Out/Out
Out/In
In/Out
In/In
2% Accuracy – Not Implemented
5% Accuracy – Not Implemented
10% Accuracy – Not Implemented
Fixed one second display update (factory
JP3-3 Alarm
Acknowledge
In
Out
Manual Acknowledge (factory)
Automatic
JP3-4 Fail
Alarm
In
Out
Enable No Counts Fail Alarm (factory)
Inhibit No Counts Fail Alarm
JP3-5
Check Source Alarm
In
Out
Alarm Inhibited (factory)
Alarm Enabled
JP4
Shield Polarity Selection
1-2
2-3
Negative Input Pulse (Other Detectors)
Positive Input Pulse (G-M Detectors)
(factory)
JP5
Shield Polarity Selection
1-2
2-3
Shield for Negative Pulse
Shield for Positive Pulse (G-M) (factory)
JP-6
Anti-Jam Fuse Selection
1-2
2-3
Enable for Normal Operation (factory)
Anti-Jam Circuit fuse bypassed (testing)
JP7
Detector Type for Anti-Jam
Circuit Timing
1-2
2-3
Out
Scintillation
G-M Tube (factory)
Anti-Jam Circuit Disabled (for testing only)