General specifications, Performance specifications – Fluke Biomedical IDA-5 User Manual
Page 43

Infusion Device Analyzer
General Specifications
33
General Specifications
Operating Voltage Range ..................................... 100 V ac to 240 V ac
Supply Frequency ................................................. 50/60 Hz
Supply Power ........................................................ <50 VA
Fuses ...................................................................... 20 mm T1.6 A H 250 V x 2
Size (H x W x D) ..................................................... 30 cm x 20 cm x 20 cm (12 in x 8 in x 8 in)
Weight .................................................................... ~5 kg (11 lb)
Altitude ................................................................... 0 m to 3000 m (10000 ft)
Temperature
Operating ............................................................ 15
°
C to 30
°
C (59
°
F to 86
°
F)
Storage ............................................................... -20
°
C to +40
°
C (-4
°
F to +104
°
F) when drained of all liquid
Humidity ................................................................. 10 % to 90 % non-condensing
Templates .............................................................. Predetermined Test Sequences. Typical capacity 200
Safety ..................................................................... IEC 61010-1: Overvoltage category II, Pollution Degree 2
Electromagnetic Environment ............................. IEC 61326-1: Portable
Emissions Classification...................................... IEC CISPR 1: Group 1, Class A. (Group 1 have intentionally generated
and/or use conductively coupled radio-frequency energy which is
necessary for the internal functioning of the equipment itself. Class A
equipment is suitable for use in non-domestic locations and/or directly
connected to a low-voltage power supply network.)
FCC ......................................................................... CFR47 Part 15.107 and 15.109 Class A
Storage of Results ................................................ Test results stored for later viewing, printing or transfer to PC. Typical
capacity: 250 tests.
Power down ........................................................... The results of tests in progress will be saved in accidental power down
Computer Control ................................................. The Product can be fully controlled using PC HydroGraph and Ansur
software for the Product
Performance Specifications
Flow Rate Measurement
Method ................................................................ Flow is calculated by measuring volume over time
Range ................................................................. 0.1 ml/h to 1500 ml/h (2500 ml/h is shown)
Accuracy ............................................................. 1 % of reading
±
1 LSD for flows of 16 ml/h to 200 ml/h for volumes
over 20 ml, otherwise 2 % of reading
±
1 LSD for volumes over 10 ml
under laboratory conditions. Degassed water at 15
°
C to 30
°
C (59
°
F
to 86
°
F) is recommended for long tests.
Max Test Duration .............................................. 100 hours
Volume Measurement
Method ................................................................ Volume is measured directly by the measuring module in minimum
sample sizes of 60
μ
l
Range ................................................................. 0.06 ml to 9999 ml
Accuracy ............................................................. 1 % of reading
±
1 LSD for flow rates of 16 to 200 ml/h for volumes over
20 ml. otherwise 2 % of reading
±
1 LSD for volumes over 10 ml under
laboratory conditions.
Max Test Duration .............................................. 100 hours
PCA Bolus/Dual Flow Measurement
Method ................................................................ See Volume measurement above
Min Bolus Volume ............................................... 0.5 ml
Resolution ........................................................... 60
μ
l increments
Max Test Duration .............................................. 100 hours
Pressure Measurement
Method (back Pressure and Flow test) ............... Direct measurement of pressure at the inlet port.
Range ................................................................. 0 psi to 45 psi or equivalent in mmHg and kPa
Accuracy ............................................................. 1 % of Full Scale
±
1 LSD under laboratory conditions
Max Test Duration .............................................. 1 hour