Calibration, Edit templates – Fluke Biomedical IDA-5 User Manual
Page 38

IDA-5
Users Manual
28
The Product can automatically do tests in a specific manner when specified
conditions are met. Test-stop preferences specifies the default actions that are to
occur. Default values can be changed at the start of each test.
Set which step to edit from the drop-down list in the When field:
•
Volume
When the set volume is reached. Zero disables
this.
•
Time
When the set time is reached. 00:00 disables this.
•
First Condition
Whichever of volume or time is first reached.
•
Both Conditions
When both the time and volume are reached.
The action when the condition is met is selected from the drop-down list
presented by the Reached then field:
•
No Action
The test continues. The user is not told.
•
Notify Operator
A beep sounds and a message shows.
•
Stop Test Only
The test stops, a beep sounds, and a message
shows.
•
Stop Test & Save
The test stops and is saved. A message tells you
the test is saved.
The Preferred Parameters let you choose default pressure units and the
position of tolerance lines on flow graphs. Psi, kPa, and mmHg pressure units
are available. The percentage tolerance can be from 0 % to 50 %.
Calibration
The Calibration menu gives access to calibration history and factory functions.
Some items on this menu are available to authorized-service personnel only.
Each screen has instructions on usage. “View Optics” is a diagnostic utility.
Edit Templates
See the “Templates” section earlier in this manual.