How to test infusion devices, Infusion device analyzer – Fluke Biomedical IDA-5 User Manual
Page 21

Infusion Device Analyzer
Product Operation
11
How to Test Infusion Devices
When you test an infusion device, it is recommended that you record the details
of the device you are testing and the test parameters. From the Status All
Channels screen, select SETUP for the relevant channel. These examples use
channel 1.
The Channel 1 Setup screen described in Table 4 is shown.
Table 4. Channel Setup Screen Controls
C h a n n e l 1 S e t u p
Device Information
Template
Occlusion
Flow
PCA/Dual Flow
Press ESC for Status Screen
gir06.eps
Control Description
Device Information
Shows a screen that lets you record infusion device data and test details.
Template
Shows a screen that lets you set a template to control a sequence of tests.
Occlusion
Occlusion test without details.
Flow
Flow test without details.
PCA/Dual Flow
PCA or dual flow test without details – asks for necessary PCA/dual flow
parameters.