Fluke Biomedical IDA-5 User Manual
Page 32

IDA-5
Users Manual
22
When ENTER is pushed in the Loading Dose field, the PCA screen changes to
the Prime Mode screen shown in Figure 8. Refer to the Flow Tests section and
prime the channel.
gir18.bmp
Figure 8. PCA Screen - Prime Mode
When the channel is primed, the Prime button is replaced by the AutoStart and
Start buttons. If the basal flow rate is 0.00, select Start to begin the test. Select
Autostart if the basal flow rate is not 0.00.
When the test is in operation, the product monitors basal flow rate, number of
boluses delivered, bolus flow rates, volumes, and durations as shown in
Table 16. Bolus intervals are also monitored. Figure 9 shows a graph of the
supplied boluses and average parameters for all boluses. When a dual flow
pump is tested, it is shown as a single, large bolus followed by the secondary
flow.
Table 16. PCA Screen in Active Mode
gir19.bmp
Menu Control
Description
Graph
Shows a graph of the test. See Figure 9.
Trig
Adds a marker to the saved data when pushed. Push ENTER at the same time as
patient demand on the pump.
End
Stops the test.