Test fluid, Storage, Shipping – Fluke Biomedical IDA-5 User Manual
Page 42

IDA-5
Users Manual
32
Test Fluid
The Product is intended to operate with de-ionized water with added detergent.
Fluids intended for use on patients, such as high viscosity, oily, or corrosive
substances will damage the measurement system. Tap water can contain
contaminates which can also damage the transducer.
Test fluid can be made with de-ionized water and a wetting agent such as
MICRO-90. It is recommended that a 0.1 % solution of MICRO-90 in de-ionized
water (preferably degassed) be prepared in volume for daily use and kept in a
sealed vessel. If the water makes too much foam, then a 0.05 % dilution is
recommended.
MICRO-90 is available from:
International Product Corp.
201 Connecticut Dr.
P.O. Box 70
Burlington, NJ 08016-0070 USA
Tel 609 386 8770
International Product Corp.
1 Church Row
Chistlehurst, Kent BR7 5PG United Kingdom
Tel. 0208 467 8944
Storage
Remove all water from the Product before storage, particularly if temperatures
can fall below 5
°
C (41
°
F). Do not pressurize the inlet ports. It is safest to use a
medical suction pump to drain the measuring channels and use the Cycle Valves
from the Calibration menu (follow on-screen instructions).
Shipping
Remove all liquid from the Product before shipping. To prevent liquid from
entering the ports, put the Product in a large plastic bag. Put the bagged Product
into its shipping carton. If this is not available, make sure there is shock
protection with a minimum of 5 cm compressible cushioning inside the carton (for
example, 60 cm x 60 cm x 60 cm).