Template operation – Fluke Biomedical IDA-5 User Manual
Page 35

Infusion Device Analyzer
Templates
25
5. Repeat for steps 2 and 3. Use the values shown in Table 17.
6. At step 4, keep the Type field empty and push ENTER. Save is highlighted.
7. Push ENTER. The Templates screen is shown with your template added to
the list.
8. Select ESC to access the Utilities menu.
9. Select ESC again to access the Status All Channels screen.
Template Operation
To use a Template:
1. From the Status All Channels screen, push SETUP for the channel to be
used.
2. Select Template.
3. Push the up and down arrow keys to select the correct template.
4. Push ENTER. Start is highlighted.
5. Push ENTER. The Device Information screen in Table 5 shows with values
from Step 1 of the template.
6. Record a control number and operator. The values from the template are
skipped.
7. Select More if it is necessary to record more data about the infusion device or
select the test type defined in the template.
8. The test operates until the set volume or time is reached. It automatically
goes to the Flow Terminated screen with a message “Set Volume Reached”
or “Set Time Reached” as applicable. A new menu button, Next Template
Step, is shown and highlighted.
9. Push ENTER on the Next Template Step button and you are prompted to
set the infusion device for the subsequent step of the template.
10. Do each template step as prompted. After the last step, Save is highlighted
on the Test Terminated screen.
11. Select the Save button.
12. Push ENTER. You are prompted for comments.