Average flow 59.70, Volume 20.00 – Fluke Biomedical IDA-5 User Manual
Page 27

Infusion Device Analyzer
Flow Tests
17
Table 11. Flow Screen in End Mode
C ha nn el 1 F l o w
Average Flow 59.70
ml/h
Volume 20.00
ml
Elapsed Time
00:20:00
Inst. Flow
0.00
ml/h
Back Pressure
0
mmHg
Save
Save & Print
Graph
Delete
Press ESC for Status Screen
gir12.eps
Menu Control
Description
Save
Keeps the results of the test after prompt for test data and comments.
Save & Print
Keeps the results and prints the results.
Prints the test results. Does not save the results.
Graph
Shows a graph of the test.
Delete
Erases the results after confirmation.
Caution
To avoid inaccurate readings, always repeat a test when
“Bubble” or “Air Lock” is shown on the display while a test is
running. See the “Troubleshooting” section of this manual.