Product maintenance, Clean the product, Outside – Fluke Biomedical IDA-5 User Manual
Page 40: Inside

IDA-5
Users Manual
30
Product Maintenance
The subsequent sections tell you how to maintain the Product.
Clean the Product
Warning
To prevent possible electrical shock, fire, or personal injury:
•
Turn the Product off and remove the mains power cord.
Stop for two minutes to let the power assemblies discharge
before you open the fuse door.
•
Do not operate the Product with covers removed or the case
open. Hazardous voltage exposure is possible.
•
Disconnect the mains power cord before you remove the
Product covers.
•
Remove the input signals before you clean the Product.
•
Use only specified replacement parts.
•
Use only specified replacement fuses.
•
Have an approved technician repair the Product.
•
Do not pour fluid onto the Product surface; fluid seepage
into the electrical circuitry may cause the Product to fail.
•
Do not use spray cleaners on the Product; such action may
force cleaning fluid into the Product and damage electronic
components.
Caution
To prevent possible damage to the Product or to equipment
under test, remove the input signals before you clean the
Product.
After troubleshooting or maintenance, restart the Product and ensure it starts
without errors. (See Product Operation.)
Outside
To clean the outside of the Product, disconnect from the power supply and use
only a damp cloth with mild detergent.
Inside
It is possible that microbial growth can become present in the transducers of the
Product. It is recommended that you clean the transducers at 3 month intervals.
To clean the inside of the Product, inject 20 ml of a warm water and detergent
solution into the Fluid Inlet Port. After 5 minutes, flush with clean water. Always
pass water from the fluid inlets to the outlets.